Determination of the drug content of pharmaceuticals containing phenothiazine compounds by triangle programmed flow titration |
| |
Authors: | Zsófia Fehér Ilona Kolbe and Ern Pungor |
| |
Institution: | (1) Technical Analytical Research Group of the Hungarian Academy of Sciences, Institute for General and Analytical Chemistry, Technical University of Budapest, Gellért tér 4, H-1111 Budapest, Hungary |
| |
Abstract: | Summary A flow-through triangle programmed coulometric titration technique is worked out for the determination of phenothiazine compounds. The oxidation reaction of phenothiazines with hypobromite ion served as titration reaction. A biamperometric detector cell has been used for following the titration process. The titration technique has been used for the determination of the drug content (average and individual) of chlorpromazine, promethazine, diethazine and trifluoperazine in pharmaceutical preparations. The relative standard deviation was in the range of 2.0 to 3.5%.
Bestimmung des Wirkstoffgehaltes von Phenothiazin-haltigen Pharmazeutica mit Hilfe einer Dreieck-programmierten Durchflu\-Titration |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|