首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Improving mass spectrometer sensitivity using a high-pressure electrodynamic ion funnel interface
Authors:Yehia Ibrahim  Keqi Tang  Aleksey V Tolmachev  Alexandre A Shvartsburg  Richard D Smith
Institution:Biological Sciences Division, Pacific Northwest National Laboratory, Richland, Washington 99352, USA.
Abstract:We report on a new electrodynamic ion funnel that operates at a pressure of 30 torr with no loss of ion transmission. The enhanced performance compared with previous ion funnel designs optimized for pressures of <5 torr was achieved by reducing the ion funnel capacitance and increasing the RF drive frequency (1.7 MHz) and amplitude (100-170 V peak-to-peak). No degradation of ion transmission was observed for pressures from 2 to 30 torr. The ability to operate at higher pressure enabled a new tandem ion funnel mass spectrometer interface design that can accommodate a greater gas load (e.g., from an ESI source). When combined with a multicapillary inlet, the interface provided more efficient introduction of ions, resulting in a significant enhancement in mass spectrometer sensitivity and detection limits.
Keywords:
本文献已被 ScienceDirect PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号