Comparison of the performances of a particle beam LC-MS interface obtained by standard and modified skimmers |
| |
Authors: | Magi Emanuele Di Carro Marina |
| |
Institution: | Dipartimento di Chimica e Chimica Industriale, Sezione di Chimica Analitica ed Ambientale, Università di Genova, Via Dodecaneso 31, 16146 Genova (I). magie@chimica.unige.it |
| |
Abstract: | Stainless steel skimmers were made in the technical workshop of our Department and mounted on a commercial Particle Beam interface. Their performances were compared with the standard ones, using three different compounds: benzidine, caffeine and perylene. Library searchable EI spectra were collected with a high quality match and remarkable improvements in the sensitivity were obtained (detection limit of 20 pg for perylene and benzidine), probably due to their better tightness that allowed to widen the skimmer orifices without significantly compromising the ion source vacuum. Also calibration curves showed a good linearity, in the concentration range 10-200 ng/ml. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|