首页 | 本学科首页   官方微博 | 高级检索  
     


Bioanalytical device based on cholesterol oxidase-bonded SAM-modified electrodes
Authors:A. Parra  E. Casero  F. Pariente  L. Vázquez  E. Lorenzo
Affiliation:(1) Departamento de Química Analítica y Análisis Instrumental, Universidad Autónoma de Madrid, Campus de Cantoblanco, 28049 Madrid, Spain;(2) Instituto de Ciencia de Materiales de Madrid (CSIC), C/ Sor Juana Inés de la Cruz 3, 28049 Madrid, Spain
Abstract:A rapid, simple and reproducible two-step method for constructing cholesterol biosensors by covalently bonding cholesterol oxidase (ChOx) to a 3,3′-dithiodipropionic acid di(N-succinimidyl ester) (DTSP)-modified gold electrode is described. Exhaustive characterizations of both the immobilization process and the morphological properties of the resulting ChOx monolayer were performed via a quartz crystal microbalance (QCM) and atomic force microscopy (AFM) operated under liquid conditions, respectively. In addition, scanning electrochemical microscopy (SECM) measurements were performed in order to check that the immobilized enzyme retains its catalytic activity. The replacement of the natural electron acceptor (O2) in the enzymatic reaction with an artificial mediator, hydroxymethylferrocene (HMF), was also studied. Finally, cholesterol was amperometrically determined by measuring the hydrogen peroxide produced during the enzymatic reaction at +0.5 V. The optimized cholesterol biosensor exhibited a sensitivity of 54 nA mM−1 and a detection limit of 22 μM.
Keywords:Cholesterol oxidase biosensor  SAM-modified electrodes  QCM  AFM  SECM
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏