首页 | 本学科首页   官方微博 | 高级检索  
     


Fast multi-planar gradient echo MR imaging: impact of variation in pulse sequence parameters on image quality and artifacts
Authors:Li Tao  Mirowitz Scott A
Affiliation:Department of Diagnostic Radiology, University of Maryland Medical Center, 22 South Greene Street, Baltimore, MD 21201-1595, USA. tli@umm.edu
Abstract:The purpose of this investigation was to quantitatively evaluate the practical impact of alteration of key imaging parameters on image quality and artifacts in fast multi-planar gradient echo (GRE) pulse sequences. These include multi-planar GRASS (MPGR) and fast multi-planar spoiled GRASS (FMPSPGR). We developed a composite phantom with different T(1) and T(2) values comprising the range of common biological tissues, which was also subjected to periodic motion in order to evaluate motion effects. Magnetic resonance imaging was performed on a GE Signa 1.5-T system. Experimental variations in key parameters included excitation flip angle (FL), echo time (TE), repetition time (TR), and receive bandwidth (BW). Quantitative analysis consisted of signal-to-noise-ratio (SNR) and contrast (CN), image nonuniformity (NU), full-width-at-half-maximum (FWHM) (i.e., blurring or geometric distortion), and ghosting ratio (GR). We found that flip angle, TE, and TR play particularly critical roles in determining image signal, homogeneity, and ghosting artifact with these sequences. Optimum clinical application of these pulse sequences requires careful attention to these imaging parameters and to their complex interactions.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号