首页 | 本学科首页   官方微博 | 高级检索  
     


Linearization of column responses by correlation chromatography. Separation improvement and reduction of undesired adsorption effects
Authors:D. J. Louwerse  H. C. Smit
Affiliation:(1) Laboratory for Analytical Chemistry, University of Amsterdam, Nieuwe Achtergracht 166, 1018 WV Amsterdam, The Netherlands
Abstract:
Summary Experiments and simulations prove that correlation chromatography can greatly reduce the disadvantage of a non-linear response of the chromatographic column. A factor that has been accepted as being an important source of error in correlation or multiplex chromatography, has been shown not to be. Separations affected, improve dramatically when correlation chromatography is used, and a substantial amount of correlation noise only arises when there is a large difference in separation between a conventional chromatogram and a correlogram.A model has been developed for simulating these nonlinearities. It is shown that, especially for correlation chromatography, the simulation results match the practical measurements very well.
Keywords:Gas chromatography  Correlation chromatography  Simulation model  Adsorption effects  Separation improvement
本文献已被 SpringerLink 等数据库收录!