首页 | 本学科首页   官方微博 | 高级检索  
     


Rapid and sensitive determination of pertechnetate in99Mo/99mTc generator eluates by reversed-phase high-performance liquid chromatography
Authors:R. Amano
Affiliation:(1) School of Allied Medical Professions, Kanazawa University, 920 Kodatsuno, Kanazawa, (Japan)
Abstract:Reversed-phase high-performance liquid chromatography with u.v. detection was applied for rapid and sensitive determination of pertechnetate in99Mo/99mTc generator eluates, using a mixture solvent of acetonitrile and 0.04M aqueous acetate buffer (1/1) containing a few volume percentage of 0.5 M tetra-n-butylammonium hydroxide as the mobile phase. Employing a mgr-bondapak C13 column, the TcO4 species was separated, monitored with absorbance at 254 nm, and observed at the retention time of 3.5 min. The detection limit was found to be 5.2·10–10 g of Tc for each injection. Total Tc contents in the99mTc eluates from clinically-used99Mo/99mTc generator were analyzed by this technique. The99mTc (99Tc) species was separated from the contaminant99Mo. This method was found to be useful for the purification of99mTc (99Tc) as well as the determination of total Tc content.
Keywords:
本文献已被 SpringerLink 等数据库收录!