首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Nitron tetrachloroaurate(III) electrodes with poly(vinyl chloride) and liquid membranes for the selective determination of gold
Authors:Hassan S S  Elnemma E M  Hamada M A
Institution:Department of Chemistry, Faculty of Science, Ain Shams University, Cairo, Egypt.
Abstract:PVC matrix and liquid membrane electrodes have been developed for direct potentiometric determination of gold(III). The membranes incorporate nitron tetrachloroaurate(III) as electroactive material. Fast response for gold(III) over the concentration range 10(-5)-0.1M, with response slopes of 52.8-55.2 mV/decade is obtained. The electrodes show good selectivity for gold(III) at pH 2-5 in the presence of many anions and cations. The PVC membrane electrode offers the advantages of greater selectivity (except for Cr(3+), Mn(2+) and ClO(-)(4)) and higher thermal stability. The liquid membrane electrode gives a higher response slope and faster time of response than the PVC membrane electrode. Determination of AuCl(-)(4) over the range 2 mug/ml-2 mg/ml shows an average recovery of 98.5% and a mean standard deviation of 1.0%. Determination of gold in some gold alloys (58.3-99.9% Au) and pharmaceutical preparations gave an average recovery of 99.4% and a mean standard deviation of 0.7%, which are comparable with the performance obtained with the spectrophotometric Malachite Green and gravimetric U.S. Pharmacopeia methods.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏