Spectrophotometric determination of selenium(IV) with 5,5-dimethyl-1,3-cyclohexanedione |
| |
Authors: | Bodini M E Pardo J Arancibia V |
| |
Affiliation: | Facultad de Química, Pontificia Universidad Católica de Chile, Casilla 6177, Santiago, Chile. |
| |
Abstract: | 5,5-Dimethyl-1,3-cyclohexanedione (dimedone) reacts in acid aqueous solution with selenium(IV) to give a benzoxaselenol which has an absorption maximum at 313 nm with a molar absorptivity of 4.00 x 10(3) l.mole(-1).cm(-1). The compound is extractable into chloroform, to give a solution with an absorption maximum at 300 nm with a molar absorptivity of 3.77 x 10(3) l.mole(-1).cm(-1). The calibration graph is linear up to 30 ppm selenium, with a detection limit of 0.1 ppm in the final solutions. Of the various other ions tested, only iron(III) interferes at all concentrations but the addition of 1000 ppm fluoride will mask 50 ppm Fe(3+). The method has good reproducibility, with a relative standard deviation of 1.0% for pure solutions. The method has been applied to the analysis of fire-refined copper. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|