首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of mercury in environmental samples by cold vapour generation and atomic-absorption spectrometry with a gold-coated graphite furnace
Authors:Soo H L  Kyung-Hoon J  Dong S L
Affiliation:Department of Chemistry, Korea Advanced Institute of Science and Technology, P.O. Box 150 Chongyang, Seoul 130-650, Korea.
Abstract:Mercury is determined at below the pg/ml level by a combination of cold vapour generation, trapping in a gold-coated graphite furnace and atomic-absorption detection. The mercury is reduced to the element by stannous chloride, stripped from solution by a stream of nitrogen and collected on a gold-coated porous graphite disk in a graphite furnace. It is then atomized by increasing the graphite furnace temperature and detected by an atomic-absorption spectrophotometer. The absolute detection limit and the characteristic mass were found to be 5 and 20 pg for 0.0044 absorbance, respectively. The concentration limit of detection was 0.1 pg/ml for a 50-ml sample, and the linear dynamic range covered three orders of magnitude. The precisions of the measurements were 2.7% for 0.1 ng and 2.6% for 2 ng of mercury. Analyses of NBS and NIES reference materials showed quantitative recovery. Analytical results obtained by the technique are presented for natural waters, marine biota and sediments.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号