Determination of mannitol and sorbitol in infusion solutions by capillary zone electrophoresis using on-column complexation with borate and indirect spectrophotometric detection
Authors:Pospísilová Marie  Polásek Miroslav  Safra Jirí  Petriska Ivan
Affiliation:Department of Analytical Chemistry, Faculty of Pharmacy, Charles University in Prague, Heyrovského 1203, CZ-500 05 Hradec Králové, Czech Republic. marie.pospisilova@faf.cuni.cz
Abstract:Capillary zone electrophoresis with indirect UV detection at 215 nm was applied for the separation and determination of mannitol (MA), sorbitol (SO) and xylitol in the form of anionic borate-polyol complexes. The separation was carried out in a fused silica capillary (total length 60 cm, effective length 50 cm, I.D. 50 microm) at 25 kV. The optimized background electrolyte was 200 mM borate buffer (pH 9.3, adjusted with triethylamine) containing 10 mM 3-nitrobenzoate as the chromogenic co-ion. The separation took approximately 13 min. The rectilinear calibration range was 0.2-2 mg mL(-1) for MA and SO when using xylitol (1 mg mL(-1)) as the internal standard. The limit of detection at a S/N of 3 was approximately 30 microg mL(-1) for either analyte. The method was used for the assay of MA or SO in pharmaceutical infusion solutions. The RSD values were 0.15% or 1.07% (n=6) when determining 100 mg mL(-1) of MA or 50 mg mL(-1) of SO in commercial infusion solutions. The results were in good agreement with those of pharmacopoeial iodimetric titration.
Keywords:
本文献已被 PubMed 等数据库收录!