首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Chemometric analysis of high performance liquid chromatography-diode array detection-electrospray mass spectrometry of 2- and 3-hydroxypyridine
Authors:Dunkerley S  Crosby J  Brereton R G  Zissis K D  Escott R E
Institution:School of Chemistry, University of Bristol, Cantock's Close, UK.
Abstract:Triply coupled high performance liquid chromatography using diode array detection and positive ion electrospray mass spectrometry of 2- and 3-hydroxypyridine is presented. Considerations of the physical method for coupling the two detectors, the influence of pH on retention times, the cone voltage of the mass spectrometer and the linear concentration ranges are described. Data from both detectors are aligned and interpolated. The analyte mass spectra are reduced to 20 significant masses. Principal components plots on the raw, normalised and standardised data, derivatives to determine composition 1 regions, deconvolution and procrustes analysis to compare data from both detectors are discussed. Common trends in both mass spectral and diode array chromatograms are interpreted. This paper represents a new approach to common processing of chromatographic data from two detectors.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏