首页 | 本学科首页   官方微博 | 高级检索  
     


A Miniature Electrolytic Conductivity Probe with a Wide Linear Range
Authors:Sang‐Ryoul Park  Harold Swerdlow
Affiliation:1. Health Metrology Group, Korea Research Institute of Standards and Science, 1 Doryong‐dong, Yuseong‐gu, Daejeon 305‐600, South Korea;2. The Dolomite Centre Ltd., 1 Anglian Business Park, Orchard Road, Royston, Hertfordshire, SG8 5TW, UK
Abstract:A novel miniature electrolytic conductivity probe and its successful operation with modified bipolar pulse conductometry are presented. The probe based on a concentric ring‐disk electrode configuration of less than 1 mm in diameter featured extremely small detection volume, and conductivity was measured with a 1 μL solution. Rapid response, easy and quick washing, and virtually no consumption of samples by measurements were additional advantages of the suggested probe design. Poor linearity was observed with conventional AC conductometry due to a large cell constant and a small double layer capacitance of the probe. Measurement circuits for modified bipolar pulse conductometry were designed, and optimization of the various pulse parameters led to a wide linear dynamic range of 0.05–10 mS cm?1, thus achieving high accuracy with a single‐point calibration. The suggested conductometric device could be conveniently applied to biological reagents and samples that are usually too little to be measured with conventional conductometers.
Keywords:Miniature sensor  Electrolytic conductivity probe  Concentric ring‐disk electrodes  Bipolar pulse conductometry
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号