Standardized method comparison for ACS:180 plus and Immulite sensitive PSA (sPSA) measurement methods |
| |
Authors: | Yusuf Kurtulmus T. Tanyalçın Giray Bozkaya Ogˇuzhan Gündüz Ömür Çerçi Fatma Z. Kutay Biltan Ersöz |
| |
Affiliation: | Ege University Faculty of Medicine, Department of Biochemistry Bornova 35100 Izmir Turkey e-mail: tanyalci@med.ege.edu.tr Tel: +90-232-388 10 96 Fax: +90-232-373 94 77, TR SSK Bozyaka Hospital, Department of Biochemistry, Izmir, Turkey, TR Ege University Faculty of Medicine, Department of Urology, Izmir, Turkey, TR Atatürk State Hospital, Izmir, Turkey, TR
|
| |
Abstract: | The aim of this study was to compare two immunoassay measurement methods, the ACS:180 Plus (Chiron Diagnostics) kit and the Immulite sPSA (DPC) kit. Method comparison analysis was performed according to EP9-A; approved guideline of NCCLS 51. Serum samples having a wide range of total prostate-specific antigen (PSA) concentrations were evaluated in split-sample analysis. F-test, t-test analysis and regression statistics were performed. In Deming regression analysis the coefficients were as follows; the slope=0.967; y-intercept=-0.148, r=0.989. An acceptable bias was seen since the systematical error was calculated to have a value less than the total allowable error calculated from biological variations. Non-parametric evaluation of the area under ROC curves for ACS:180 Plus and Immulite sPSA were 0.997 and 0.987, respectively. Diagnostic accuracy was at the level of p= 0.000 and no statistical difference was found between the two assay methods. |
| |