首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Temperature Calibration in Dielectric Measurements
Authors:J F Mano  S Lanceros-Méndez  A M Nunes  M Dionísio
Institution:(1) Departamento de Engenharia de Polímeros, Universidade do Minho, Campus de Azurém, 4800-058 Guimarães, Portugal;(2) Departamento de Física, Universidade do Minho, Campus de Gualtar, 4710 Braga Codex, Portugal;(3) Departamento de Química, Centro de Química Fina e Biotecnologia, Faculdade de Cieogonncias e Tecnologia, Universidade Nova de Lisboa, 2825-114 Monte de Caparica, Portugal
Abstract:Dielectric experiments are often performed in non-isothermal conditions. Thus, there is a difference between the temperature of the sample and the sensor temperature. In this work we propose and compare three temperature calibration methods based on the detection of transitions or relaxations: i) the melting of high-purity metallic standards (indium and tin), ii) the 2nd order phase transition of a ferroelectric crystal (TGS); iii) the agr-relaxation of an amorphous polymer (poly(carbonate)). The results obtained from the three different methods were used to construct a calibration curve for a given heating rate.This revised version was published online in November 2005 with corrections to the Cover Date.
Keywords:dielectric  metallic standards  polymer  temperature correction  TGS  TSC