首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Simultaneous quantification of reduced and oxidized glutathione in plasma using a two-dimensional chromatographic system with parallel porous graphitized carbon columns coupled with fluorescence and coulometric electrochemical detection
Authors:Sakhi Amrit K  Russnes Kjell Magne  Smeland Sigbjørn  Blomhoff Rune  Gundersen Thomas E
Institution:Institute of Basic Medical Sciences, Department of Nutrition, University of Oslo, P.O. Box 1046 Blindern, N-0316 Oslo, Norway.
Abstract:A method for the simultaneous quantification of reduced and oxidized glutathione in human plasma employing a two-dimensional chromatographic system with parallel porous graphitized carbon (PGC) columns coupled with fluorescence (FLD) and coulometric electrochemical detection (ED) has been developed. Post-sampling oxidation of reduced glutathione (GSH) was prevented by derivatizing the -SH group with monobromobimane (MBB) and the glutathione-bimane adduct (GSMB) was detected by FLD. Oxidized glutathione (GSSG) was detected by ED optimized to give lowest possible limits of detection (LOD). The method is fully validated and is currently used for determination of GSH, GSSG and its redox potential in different clinical studies.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏