Preparation of Pt/polypyrrole-ferrocene hydrogen peroxide sensitive electrode for the use as a biosensor |
| |
Authors: | F Arslan A Yaşar E Kiliç |
| |
Institution: | (1) Gazi University, 06500 Ankara, Turkey;(2) Ankara University, Ankara, Turkey |
| |
Abstract: | A polypyrrole electrode with ferrocene mediator is prepared and its sensitivity to hydrogen peroxide is investigated. The
polypyrrole is deposited upon a 0.5 cm2 Pt plate by the polymerization of pyrrole by scanning the electrode potential between 0.0 and 0.9 V at a scan rate of 50
mV/s. The platinum/polypyrrole-ferrocene (Pt/PPy-Fc) electrode is prepared by adding ferrocene to the coverage medium. The
electrode’s sensitivity to hydrogen peroxide is investigated at room temperature using 0.025 M phosphate buffer at pH 7. The
working potential is 0.7 V, the concentrations of pyrrole and ferrocene are 0.2 M and 10 mM. Polypyrrole was coated on the
electrode surface within 26 cycles.
Published in Russian in Elektrokhimiya, 2006, Vol. 42, No. 2, pp. 160–164.
The text was submitted by the authors in English. |
| |
Keywords: | polypyrrole hydrogen peroxide ferrocene |
本文献已被 SpringerLink 等数据库收录! |
|