首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Automated sequential injection fluorimetric determination and dissolution studies of Ergotamine Tartrate in pharmaceuticals
Authors:Legnerová Z  Sklenárová H  Solich P
Institution:

Laboratory of Flow Analysis, Department of Analytical Chemistry, Faculty of Pharmacy, Charles University, Heyrovského 1203, 500 05, Hradec Králové, Czech Republic

Abstract:A fully automated flow system for drug-dissolution studies based on the sequential injection analysis (SIA) was described and used for monitoring dissolution profiles of Ergotamine Tartrate (ET) in pharmaceutical formulations. 50 μl of dissolution medium was taken for each measurement at a flow rate of 40 μl s?1 and detected by fluorescence detector using λex=236 nm (λem≥390 nm). The calibration curve was linear over the range 0.03–0.61 mg l?1 of ET (sufficient for the dissolution tests). Equation of the calibration curve was calculated giving the following values: F=117.7 c+0.80 (n=6); r=0.9998. Detection limit was 0.01 mg l?1 of ET. The R.S.D. is less than 0.54 and 0.86% (n=10) when determining 0.61 and 0.03 mg l?1 of ET in standard solution, respectively. The dissolution test of Bellaspon tablets (0.3 mg of ET in 1 tablet) was programmed for 20 min, with a continuous sampling rate of 120 h?1 under conditions required by BP 1993.
Keywords:Sequential injection analysis (SIA)  Dissolution test  Ergotamine Tartrate  Fluorimetric detection
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号