首页 | 本学科首页   官方微博 | 高级检索  
     


Application of amperometric sol-gel biosensor to flow injection determination of glucose
Authors:Couto Cristina M C M  Araújo Alberto N  Montenegro M Conceição B S M  Rohwedder Jarbas  Raimundo Ivo  Pasquini Célio
Affiliation:

a CEQUP/Departamento de Química-Física, Faculdade de Farmácia (UP), Rua Aníbal Cunha, 164-4050 Porto, Portugal

b Instituto de Química, Universidade Estadual de Campinas, C.P 6145 Campinas, Brazil

c Instituto Superior de Ciências de Saúde, Norte, Rua Central da Gandra, 1317–4580 Gandra, Paredes, Portugal.

Abstract:A glucose biosensor with enzyme immobilised by sol–gel technology was constructed and evaluated. The glucose biosensor reported is based on encapsulated GOX within a sol–gel glass, prepared with 3-aminopropyltriethoxy silane, 2-(3,4-epoxycyclohexyl)-ethyltrimetoxy silane and HCl. A flow system incorporating the amperometric biosensor constructed was developed for the determination of glucose in the 1×10−4–5×10−3 mol l−1 range with a precision of 1.5%. The results obtained for the analysis of electrolytic solution for iv administration and human serum samples showed good agreement between the proposed method and the reference procedure, with relative error <5%.
Keywords:Flow injection analysis   Glucose oxidase   Sol–gel   Glucose biosensor