Optimization of a peak compression system for a remoxipride metabolite (FLA797) and its application to bioanalysis. |
| |
Authors: | L B Nilsson |
| |
Affiliation: | Bioanalysis, Astra Arcus AB, S?dert?lje, Sweden. |
| |
Abstract: | A peak compression system is optimized for FLA797 (I), a phenolic tertiary amine and a metabolite to the antipsychotic drug remoxipride. An application is described where this effect is used to give a 6-7-fold improvement of the quantification limit in an assay of I in plasma. The liquid chromatographic system is constructed so that the injection of I dissolved in a solution of a competing amine gives a very high and narrow analyte peak with an apparent efficiency of 1.5 x 10(6) plates/m. When the levels of I in plasma are determined, an internal standard, giving a normal isocratic peak, is added to the plasma sample before the extraction. Concentrations of I down to 0.5-1.0 nM can be determined with reasonable precision. FLA908, another phenolic remoxipride metabolite and a regioisomer to I, eluting as a normal isocratic peak, can be determined simultaneously although only at concentrations higher than 10-15 nM. |
| |
Keywords: | |
|
|