A mid-infrared spectrometer based on difference frequency generation
Authors:Lunhua Deng   Ling Han   Wanguo Liang   Zhensong Cao   Changqing Xu   Weijun Zhang   Zhiben Gong  Xiaoming Gao  
Affiliation:aLaboratory of Atmospheric Optics, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China;bEnvironmental Spectroscopy Laboratory, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China;cDepartment of Engineering Physics, MC Master University, 1280 Main Street West, Hamilton, Ontario, Canada
Abstract:A mid-infrared spectrometer operating in the range from 3.42 to View the MathML source based on difference-frequency mixing of a tunable Ti:Sapphire laser and a diode-pumped monolithic Nd:YAG laser in periodically poled LiNbO3 (PPLN) crystals was developed. A system line width of about 36 MHz was obtained. An energy conversion efficiency of 0.6 mW/W2 was obtained at View the MathML source. The direct absorption spectra of pure methane and pure carbon dioxide samples were measured to test the performance of the spectrometer.
Keywords:Tunable Ti:Sapphire laser   Nd-YAG laser   Trace gas detection