首页 | 本学科首页   官方微博 | 高级检索  
     


Reliability and performances of a high-purity gamma spectrometry system used for environmental measurements
Authors:Ileana Radulescu  Marian Romeo Calin
Affiliation:1. “Horia Hulubei” National Institute for Physics and Nuclear Engineering (IFIN-HH), 30 Reactorului Street, P.O. Box MG-6, 077125, Bucharest-Magurele, Romania
Abstract:Proficiency testing is among the most methods to assess the accuracy of the analytical data produced by laboratories. The results presented in this paper show the performance of the laboratory for three intercomparisons organized by Radionuclide Metrology Laboratory (RML) and by IAEA, in which SALMROM laboratory have participated in three consecutive years. The feedbacks from the final reports from RML and IAEA demonstrate the performance of the laboratory where 34 out of 39 pass all acceptance criteria of the test. The rest of five results which failed served to improve the laboratory analysis and to optimize accuracy measurements. The overall results of the proficiency tests point out the reliability and traceability of the gamma spectrometry system used for environmental measurements.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号