Needle‐trap device for the sampling and determination of chlorinated volatile compounds
Authors:Elżbieta Dobrzyńska  Bogusław Buszewski
Institution:1. Department of Chemical, Aerosol and Biological Hazards, Central Institute for Labour Protection—National Research Institute, , Warsaw, Poland;2. Department of Environmental Chemistry and Bioanalytics, Faculty of Chemistry, Nicolaus Copernicus University, , Toruń, Poland
Abstract:A needle‐trap device, with immobilized sorbent inside the syringe, coupled with GC–MS was applied for air sampling and determination of chlorinated volatile organic compounds such as dichloromethane, trichloromethane, and tetrachloromethane. The application of a needle trap packed with combination of three sorbents including Tenax TA, Carbopack X, and Carboxen 1000 resulted in detection limits of few pg for chlorinated volatile compounds and recoveries of 99.2–102.8%. The extraction and desorption parameters were optimized within the study. As a result, the precision determined as RSD was equal to 5.05 and 3.03 and 6.52% for dichloromethane, trichloromethane, and tetrachloromethane, respectively. The storage time for chlorinated compounds up to 48 h and reusability of the needle‐trap device were verified. The obtained results have proved the ability of needle traps to compete with other solventless sampling and sample preparation extraction techniques.
Keywords:Chlorinated compounds  Carbopack X  Carboxen 1000  Needle trap  Tenax TA