首页 | 本学科首页   官方微博 | 高级检索  
     


A modified apparatus for determination of carbon in metals by the low pressure method
Authors:V.T. Athavale  S.P. Awasthi  N. Krishnamachari  M. Sundaresan  M.S. Varde
Affiliation:Analytical Division, Atomic Energy Establishment Trombay, Bombay -28 India
Abstract:The low pressure method of determination of carbon in metals and alloys was modified to include certain new techniques. An isopentane slush bath dispensed with the use of liquid oxygen and the usual McLeod gauge was replaced by a differential oil manometer, which increased the sensitivity. As little as 5 μg of carbon in CaCO3-quartz standards could be determined with a coefficient of variation of 16.0% which improved to 3.3% at the 100-μg level. The apparatus was used for the determination of carbon in metals such as uranium, zirconium and iron and in steels and cupronickel alloy.
Keywords:
本文献已被 ScienceDirect 等数据库收录!