首页 | 本学科首页   官方微博 | 高级检索  
     


New Validated Potentiometric Determination of Vasodilator Pentoxifylline in its Pharmaceutical Formulations and Biological Fluids
Authors:Nawal Alarfaj  Maha El‐Tohamy
Affiliation:Department of Chemistry, College of Science, King Saud University, P.O. Box 22452, Riyadh 11495, Saudi Arabia
Abstract:The construction and performance characteristics of pentoxifylline selective electrodes were developed. Two types of electrodes: plastic membrane I and coated wire II were constructed based on the incorporation of pentoxifylline with phosphotungstic acid (PTA). The influence of membrane composition, kind of plasticizer, pH of the test solution, soaking time, and foreign ions on the electrodes was investigated. The electrodes showed a Nernstain response with a mean calibration graph slope of 56.77 ± 0.19 and 55.76 ± 0.71 mV decade‐1 at 25 °C for electrode I and II respectively, over pentoxifylline concentration range from 1.0 × 10‐5‐1.0 × 10‐2 and 9.0 × 10‐6‐1.0 × 10‐2 mol L‐1, with detection limits 4.89 × 10‐6 and 3.90 × 10‐6 mol L‐1 for electrode I and II, respectively. The pH range of the constructed electrodes was 4‐6. Interferences from common cations, alkaloids, sugars, amino acids and drug excipients were reported. The results obtained by the proposed electrodes were also applied successfully to the determination of the drug in its pharmaceutical preparations and biological fluids.
Keywords:Plastic membrane  Coated wire electrode  Ion‐selective electrode  Pentoxifylline  Potentiometric determination  Pharmaceutical formulation  Biological fluids
设为首页 | 免责声明 | 关于勤云 | 加入收藏