首页 | 本学科首页   官方微博 | 高级检索  
     


Design and evaluation of an on-line microcolumn pre-concentration technique for graphite furnace atomic absorption spectrometry
Authors:E. Beinrohr, M. akrt, M. Rapta  P. Tarapí  
Affiliation:(1) Department of Analytical Chemistry, Slovak Technical University, CS-812 37 Bratislava, "Ccaron"SSR
Abstract:Summary A simple method is described for trace element pre-concentration on a microcolumn of inner diameter 1 mm and 25 mm length. The column packed with methacrylate gel with bound 8-quinolinol groups (Spheron Oxin) is incorporated into the tip of the sampling arm of the autosampler. The sample solution, water, eluent and air streams are switched with an eight-port valve. The flow-rate at pre-concentration is about 0.14 ml/min. After washing out the residues of the sample matrix from the column, the collected elements are eluted with 1 mol/l HNO3 directly into the graphite tube of the atomizer. The reproducibility of the method was found to be better than 5%. The method was used for the determination of Pb in NaCl.
Entwurf und Abschätzung einer on-line Mikrosäulen-Vorkonzentrierungstechnik für die Graphitofen-Atomabsorptionsspektrometrie
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏