Design and evaluation of an on-line microcolumn pre-concentration technique for graphite furnace atomic absorption spectrometry |
| |
Authors: | E. Beinrohr, M. akrt, M. Rapta P. Tarapí |
| |
Affiliation: | (1) Department of Analytical Chemistry, Slovak Technical University, CS-812 37 Bratislava, SSR |
| |
Abstract: | Summary A simple method is described for trace element pre-concentration on a microcolumn of inner diameter 1 mm and 25 mm length. The column packed with methacrylate gel with bound 8-quinolinol groups (Spheron Oxin) is incorporated into the tip of the sampling arm of the autosampler. The sample solution, water, eluent and air streams are switched with an eight-port valve. The flow-rate at pre-concentration is about 0.14 ml/min. After washing out the residues of the sample matrix from the column, the collected elements are eluted with 1 mol/l HNO3 directly into the graphite tube of the atomizer. The reproducibility of the method was found to be better than 5%. The method was used for the determination of Pb in NaCl.
Entwurf und Abschätzung einer on-line Mikrosäulen-Vorkonzentrierungstechnik für die Graphitofen-Atomabsorptionsspektrometrie |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|