Determination of selenium in soil by hydride generation AAS |
| |
Authors: | V Kos Marjan Veber and Vida Hudnik |
| |
Institution: | (1) Analytical Chemistry Lab, National Institute of Chemistry, Hajdrihova 19, 1000 Ljubljana, Slovenia, XX;(2) University of Ljubljana, Faculty of Chemistry & Chemical Technology, 1000 Ljubljana, Slovenia, XX |
| |
Abstract: | Hydride generation AAS was applied for the determination of total selenium in soil. The influence of various anions and cations
present in the sample solution on the generation of selenium hydride was investigated. Special attention was paid to the wet
oxidation procedure for the sample dissolution. The proposed procedure involves microwave sample preparation by using a mixture
of H2O2/HNO3/H2SO4. The accuracy of the procedure was checked by the analysis of the standard reference material Buffalo River Sediment (SRM-2704,
NIST). Satisfactory agreement of the results obtained with the values reported was achieved. The detection limit of the procedure
was 25 ng Se/g of soil. The relative standard deviation of the measurements varied from 5.5% for SRM-2704 up to 10% for the
real soil samples.