Electrochemical flow-through detector for the determination of cystine and related compounds |
| |
Authors: | R Eggli R Asper |
| |
Institution: | Institute of Inorganic Chemistry, University of Zürich, CH-8001 ZürichSwitzerland;University Hospital of Zürich, CH-8091 ZürichSwitzerland |
| |
Abstract: | An electrochemical double cell for the detection of cystine in aqueous liquid streams is described. A column electrode of amalgamated silver powder is used to reduce cystine quantitatively to cysteine, which is detected amperometrically at a mercury electrode. The double cell is applied to the detection of cystine, cysteine and penicillamine in high-pressure liquid chromatography and to the selective determination of cystine and cysteine by flow-injection analysis. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|