首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Stable isotope dilution analysis by thermal ionization mass spectrometry (II): The determination of cadmium and copper
Authors:A Broekman and J G van Raaphorst
Institution:(1) ECN Netherlands Energy Research Foundation, P.O. Box 1, NL-1755 ZG Petten, The Netherlands
Abstract:Summary The combination of stable isotope dilution analysis (SIDA) and thermal ionization mass spectrometry (TIMS) is in use for lead and uranium determination at milligram per kilogram levels for over 20 years.However, several other elements can also be determined accurately by SIDA/TIMS. In this study the determinations of cadmium and copper are described. Details of the digestion, electrochemical and ion-exchange separations and the loading of the elements on a filament are given. The advantages of the SIDA/TIMS technique are shown and illustrated with results for several certified reference materials.
Isotopenverdünnungsanalyse mit stabilen Isotopen durch Thermionen-Massenspektromctrie: Bestimmung von Cadmium und Kupfer
Zusammenfassung Die Kombination der Isotopenanalyse mit stabilen Isotopen mit der Thermionenmassenspektrometrie wird schon seit über 20 Jahren zur Bestimmung von Blei und Uran im mg/kg-Bereich angewendet. Sie ist jedoch auch zur genauen Bestimmung anderer Elemente geeignet und wird in dieser Arbeit für Cadmium und Kupfer eingesetzt. Aufschlu\, Trennungen durch elektrochemische Methoden und Ionenaustausch sowie das Aufbringen auf den Heizfaden wurden im einzelnen beschrieben. Die Vorteile des Verfahrens werden an Hand von Ergebnissen für zertifizierte Standardmaterialien aufgezeigt.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏