首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Theory of errors of the radioanalytical method of sub-super equivalence using a simple solvent extraction system
Authors:M Kyr?  K P?ikrylová
Institution:(1) Institute of Nuclear Research, 250 68 Řež, (Czechoslovakia)
Abstract:The theory is given of errors of the title method based on the solvent extraction reaction Aaq+Baq⇌ABorg, the minimum amounts which can be determined are assessed, the desirable values of ξ are recommended and the precision of the title method is compared with that of the concentration dependent distribution method. Tables are given allowing a rapid calculation of the error of determination for any reasonable concentration of the substance to be determined, any extractant concentration and a chosen value of ξ if the equilibrium constant of the above reaction is known.