首页 | 本学科首页   官方微博 | 高级检索  
     


Solid-phase extraction clean-up procedure for the analysis of PAHs in lichens
Authors:María Blasco  Celia DomeÑo  Karim Bentayeb
Affiliation:Grupo GUIA-Instituto de Investigación de Ingeniería en Aragón (I3A) , Centro Superior de Ingenieros (CPS) , María de Luna 3, 50018 Zaragoza, Spain
Abstract:A clean-up procedure based on a solid-phase extraction column was optimized for determination of polycyclic aromatic hydrocarbons (PAHs) in lichen extracts to remove co-extracted compounds from the matrix in the final extract. Several kinds of solid phases were evaluated: normal phase (-NH2 and alumina), strong anion exchange and reversed phase. The -NH2 columns were the most effective by using a packed solid bed of 500?mg. The lichen raw extract was loaded on the column previously conditioned with dichloromethane and hexane. Hexane (0.5?mL) was used as rinsing solvent, and PAHs were quantitatively eluted (80–97%) using 2?mL of hexane–dichloromethane (65–35) as eluting solvent. In these conditions, even the heaviest PAHs were quantitatively eluted. The optimized SPE method provides a short time and low-solvent-consumption sample clean-up compared with other conventional methods based on column chromatography. The analytical procedure, dynamic sonication-assisted extraction, followed by the optimized solid-phase extraction clean-up, was used to determine the 16 EPA priority PAHs from native lichens collected from the Aragon valley in central Pyrenees. The PAH concentrations in lichen samples ranged from 352 to 1654?ng?g?1, and the minimum concentration value was established as the regional reference PAH levels in the area.
Keywords:PAHs  Lichens  SPE  Clean-up
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号