首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Chromatography of Uranium on High-Performance Ion Exchangers
Authors:R M Cassidy  S Elchuk
Institution:Atomic Energy of Canada Limited Research Company, Chalk River Nuclear Laboratories , Chalk River, Ontario, Canada , KOJIJO
Abstract:Abstract

The potential of high-performance liquid chromatography (HPLC) for the determination of U(VI) in ground waters and urine has been examined under a variety of HPLC experimental conditions. Conventional cross-linked and bonded-phase ion exchangers, both cation and anion, were studied with aqueous mobile phases containing tartrate, citrate, or α-hydroxyisobutyrate. The best chromatography was obtained on bonded-phase cation exchangers with an α-hydroxyisobutyrate eluent. The metal ions were detected either by visible spectrophotometry after a post-column reaction with a complexing reagent, or with a polarographic detector. Dectection after post-column reaction gave the best sensitivity; the detection limit (2 × baseline noise was 6 ng or 60 ng.ml?1 for 100 μl samples. In-line trace enrichment was used to decrease detection limits and linear calibration curves were observed in the ranges studied; 0.5 to 50 ng.mL?1 for ground waters and 25 to 400 ng.mL?1 for artificial urine.
Keywords:Uranium  ion exchange  urine  ground water  trace analysis  
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号