Calibration and application of multi-channel soft X-ray Dante spectrometer
Zhang Siqun,Huang Xianbin,Li Jing,Yang Libing,Cui Mingqi,Zhao Yidong. Calibration and application of multi-channel soft X-ray Dante spectrometer[J]. High Power Laser and Particle Beams, 2009, 21(7)
Authors:Zhang Siqun  Huang Xianbin  Li Jing  Yang Libing  Cui Mingqi  Zhao Yidong
Affiliation:1. Institute of Fluid Physics, CAEP, P.O. Box 919-108, Mianyang 621900, China; 2. Institute of High Energy Physics, China Academy of Sciences, Beijing 100039, China
Abstract:In Z-pinch diagnostic,Dante spectrometer plays an important role in the measurement of soft X-ray spectrum and total flux.This paper reports on the structure and channel configuration of a "Dante" spectrometer designed for z-pinch experiments on Yang accelerator.Calibration of X-ray devices used in the spectrometer,such as X-ray diodes,grazing incidence mirrors and filters are described in detail and reasons accounting for the deviation from theoretical values have also been discussed.Waveforms and unfolding results which also compared with X-ray scintillator powermeter obtained in an argon gas-puff experiment are presented as well,the power of X- ray radiation is about 30~40 GW and the energy of it is about 0.8 kJ,the difference of the results of Dante spectrometer and X-ray scintillator powermeter is about 20%.
Keywords:Z-pinch  soft X-ray  Dante spectrometer  calibration