A method for measurement of free cadmium species in waters using diffusive gradients in thin films technique with an ion-imprinted sorbent |
| |
Authors: | Hong-Tao Fan Yan Lu Ai-Juan Liu Bing Jiang Hua Shen Cong-Cong Huang Wen-Xiu Li |
| |
Affiliation: | 1. Liaoning Provincial Key Laboratory of Chemical Separation Technology, Shenyang University of Chemical Technology, Shenyang, 100142, Liaoning, China;2. Food Engineering, Shandong Business Institute, Yantai, 264670, China |
| |
Abstract: | A diffusive gradients in thin films (DGT) device for the analysis of free Cd(II) species, based on Cd(II) ion-imprinted sorbent (IIS) as the binding agents and commercial polyethersulfone membrane (PES) as diffusion layer, was developed (PES/IIS-DGT). DGT time-series experiments showed that the mass of free Cd(II) species accumulated by PES/IIS-DGT was linear vs. time (R2 = 0.9953) and the concentration of free Cd(II) species by PES/IIS-DGT was in good agreement with the total dissolved concentrations of free Cd(II) species in simple synthetic solutions where free ionic species dominated. PES/IIS-DGT performance was independent in the range of pH 4.5–7.5 and ionic strength range from 1.0 × 10−3 to 0.7 mol L−1. The measurement of free Cd(II) species in synthetic solution containing different concentrations of ligands by PES/IIS-DGT showed an excellent agreement with the value measured by Cd(II) ion selective electrodes (Cd-ISE), indicating that PES/IIS-DGT method is more suitable than Cd-ISE for the measurement of low concentration of free Cd(II) species due to the enrichment of IIS for the analytes. |
| |
Keywords: | Cadmium Free species Ion-imprinted Diffusive gradients in thin-films |
本文献已被 ScienceDirect 等数据库收录! |
|