首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Optimal Parameters in High Speed Liquid Chromatography (HPLC)
Authors:Istvn Halsz  Gerhard Grlitz
Institution:István Halász,Gerhard Görlitz
Abstract:After a decade, the first storm of development in high performance liquid chromatography seems to be abating. Providing the stationary phase and eluent are optimized, most routine analytical and semi-preparative separation problems require columns having between 1000 and 10000 theoretical plates. Commercial apparatus and columns completely satisfy these requirements. For many separation problems the efficiency of the columns on offer is greater than required. Since optimal (not maximal) separation is required with shorter analysis times, the column inlet pressure is increased in order to increase the eluent velocity, while at the same time reducing the column efficiency. The same result can be obtained by using shorter columns and lower pressures. Since inlet pressures of more than 100 bar are seldom required in routine analysis, the term “high pressure liquid chromatography” is a misnomer.—It is probable that difficult separation problems, where several 100 000 theoretical plates are required, will be studied in the next phase of development. It can be shown using nomograms that such problems can be solved within acceptable time limits using inlet pressures of less than 500 bar. Further possible directions of development are indicated.
Keywords:Chromatography  HPLC (high pressure liquid chromatography)
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号