Abstract: | Negative ionization electrospray liquid chromatography mass spectrometry was developed for the simultaneous determination of all nine haloacetic acids containing bromine and chlorine. Haloacetic acids were separated on a crosslinked polystyrene resin column using 3% acetic acid dissolved in acetonitrile:water (20:80) as mobile phase. The precision of this method varied from ±2.2 to ±7.1% for nine haloacetic acids. In addition, quantitative results obtained with spiked water samples at three different concentrations are described. The limit of detection of the proposed method using 200 mL of water samples was between 0.003 and 0.070 μg/L. This method was successfully applied to the trace determination of haloacetic acids in waste water, river water, and seawater. |