首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Indirect conductimetric detection of nonelectrolytes in microcolumn liquid chromatography
Authors:T Takeuchi  T Miwa
Institution:(1) Faculty of Engineering, Gifu University, 1-1 Yanagido, 501-11 Gifu, Japan
Abstract:Summary Indirect conductimetric detection of nonelectrolytes was investigated in microcolumn liquid chromatography. The mobile phase contained an electrolytic substance which maintained a high background and analytes were indirectly detected owing to a depression of the background signal. The dynamic reserve, defined as the ratio of the background to its noise level, was 9.3×104, and the detection limits at S/N=3 were 10–32 ng for alcohols examined.
Keywords:Microcolumn liquid chromatography  Indirect conductimetric detection  Nonelectrolytes
本文献已被 SpringerLink 等数据库收录!