Implementation of the chemical reaction interface mass spectrometry technique on a Hewlett-Packard mass-selective detector
Authors:Hengchang Song  Jozef Kusmierz  Fred Abramson  Matt McLean
Affiliation:1. Department of Pharmacology, George Washington University Medical Center, 2300 I Street Northwest, 20037, Washington, DC
2. Vestee Corporation, Houston, Texas, USA
Abstract:A microwave-powered chemical reaction interface has been installed in a Hewlett-Packard gas chromatograph-mass spectrometer (GC-MS) system (5890 IT gas chromatograph-S971 mass-selective detector). The technical details and optimization strategies are discussed. The evaluation of this new setup is presented, showing detection limits of 1 ng of 13C-, 15N-, and Cl-containing compounds with signal-to-noise ratios greater than or egual to 3. Selective detection was evaluated with a urine sample from a dog dosed with 15N3-midazolam that had been previously analyzed by using a differentially pumped research-level quadrupole mass spectrometer. The results show that the detection of 15N and Cl remains highly selective and the mass-selective detector gives comparable sensitivity to the larger instrument when the latter is operating over a conventional mass range. The capability for chemical reaction interface mass spectrometry can be easily accomplished with an inexpensive GC-MS system.