首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Investigation of the effect of a dc glow discharge on the performance of a particle beam liquid chromatography/mass spectrometry interface
Authors:Thomas A Bellar  William L Budde  David D Kryak
Institution:1. Office of Research and Development, Environmental Monitoring Systems Laboratory, U.S. Environmental Protection Agency, 45268, Cincinnati, OH, USA
Abstract:A low voltage (180-V) dc glow discharge device was inserted just below the pneumatic nebulizer in a particle beam interface of a high performance liquid chromatography/mass spectrometry system. The discharge in a helium atmosphere increased the signal produced by 12 test compounds by factors of about 2–6. The increases in signal were probably produced by an increase in the efficiency of solute transmission through the interface. The signal increases caused by the glow discharge were compared to somewhat larger increases caused by 0.01-M ammonium acetate in the mobile phase. The combination of glow discharge and ammonium acetate provided no meaningful advantage over the individual techniques. The mechanism of improved transport efficiency is not proven, but the neutralization of charged particles is a viable hypothesis.
Keywords:
本文献已被 ScienceDirect SpringerLink 等数据库收录!