首页 | 本学科首页   官方微博 | 高级检索  
     


Traceable measurements of pH
Authors:Petra Spitzer
Affiliation:(1) Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany e-mail: petra.spitzer@ptb.de Tel.: +49-53-5923322 Fax: +49-53-5923015, DE
Abstract: The primary method for pH is based on the measurement of the potential difference of an electrochemical cell containing a platinum hydrogen electrode and a silver/silver chloride reference electrode, often called a Harned cell. Assumptions must be made to relate the operation of this cell to the thermodynamic definition of pH. National metrology institutes use the primary method to assign pH values to a limited number of primary standards (PS). The required comparability of pH can be ensured only if the buffers used for the calibration of pH meter-electrode assemblies are traceable to these primary pH standards. To assess the degree of equivalence, comparisons of primary measurement procedures for pH were organized in co-operation with EUROMET. Typical results will be presented. In 1998, the Consultative Committee for Amount of Substance (CCQM) decided to include the field of pH in its working programme. The first key comparison for this quantity was recently carried out on two phosphate buffer solutions. Received: 14 June 2000 / Accepted: 22 August 2000
Keywords:Metrology in chemistry  Traceability  pH  Key comparison
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号