首页 | 本学科首页   官方微博 | 高级检索  
     

Measuring normal spectral emissivities of niobium by a pulse-heating technique:1000K to the melting point
作者姓名:王青伟  萧鹏  于金涛  戴景民
作者单位:Department of Automation Measurement and Control Harbin Institute of Technology,Harbin 150001,Department of Automation Measurement and Control,Harbin Institute of Technology,Harbin 150001,Department of Automation Measurement and Control,Harbin Institute of Technology,Harbin 150001,Department of Automation Measurement and Control,Harbin Institute of Technology,Harbin 150001
基金项目:The work was supported by the National Natural Science Foundation of China (No. 60377037)the National Fund of Author of Most Excellent Doctoral Dissertation (No. 199929).
摘    要:The normal spectral emissivity of niobium strip specimen is measured using a new pulse-heating reflectometric technique. The hemispherical spectral reflectivity of the surface of a strip tangent to an integrating sphere is determined by a high-speed lock-in technique. At the same time, the radiance tem-perature of the strip is measured by the multi-wavelength high-speed pyrometer from approximately 1000 K to the melting point. Details of the measurement method and of the related calibration techniques are reported. Results of the normal spectral emissivity of niobium at 633, 753, 827, and 905 nm from room temperature to its melting point are presented. The accuracy of spectral emissivities is estimated better than 5%.

本文献已被 CNKI 等数据库收录!