首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of fumonisins B1 and B2 in corn by liquid chromatography/mass spectrometry with immunoaffinity column cleanup: single-laboratory method validation
Authors:Senyuva Hamide Z  Ozcan Sureyya  Cimen Dilek  Gilbert John
Institution:TUBITAK, Ankara Test and Analysis Laboratory (ATAL), Konya Yolu No. 67, Besevler, 06530, Ankara, Turkey. hamide.senyuva@tubitak.gov.tr
Abstract:A single-laboratory method validation was conducted to establish the effectiveness of an immunoaffinity column cleanup procedure followed by liquid chromatography/mass spectrometry (LCIMS) for the determination of fumonisins B1 and B2 (FBI + FB2) in corn. The test portion is extracted with acetonitrile-methanol-water (25 + 25 + 50). The extract is filtered, diluted with phosphate-buffered saline solution, and applied to an immunoaffinity column. FB1 + FB2 are removed with methanol and directly determined by reversed-phase LC with MS detection using selected-ion monitoring of 2 characteristic ions in each case. Test portions of blank corn samples were spiked with a mixture of FB1 + FB2 to give total levels of 200 and 500 ng/g, respectively. Recoveries of both FB1 and FB2 from spiked samples averaged 90.4-101%. Based on results for spiked raw corn (triplicates at 2 levels), the relative standard deviation for repeatability ranged from 2.8 to 7.1%. The accuracy of the method was demonstrated by analysis of Food Analysis Performance Assessment Scheme (FAPAS) test material. The method was also applied to a small survey of processed corn products such as corn chips, cornflakes, and popcorn.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号