Accurate and sensitive high-performance liquid chromatographic method for geometrical and structural photoisomers of bilirubin IX alpha using the relative molar absorptivity values.
Authors:S Itoh  K Isobe  S Onishi
Institution:Department of Pediatrics, Kagawa Medical University, Japan.
Abstract:It has been reported that considerable differences exist between the relative molar absorptivity values of the geometrical and structural photoisomers of bilirubin. We have devised an accurate HPLC method for photoisomer quantification based on the following principle: the sum of both the integrated peak areas corrected by each factor for each photoisomer, and the integrated peak area of unchanged (ZZ)-bilirubin (ZZ)-B] after an anaerobic photoirradiation, should be constant and equal to the integrated peak area of initial (ZZ)-bilirubin (ZZ)-Bi] before photoirradiation. On this basis, the following equation can be used to determine each factor. equation: see text] alpha, beta, gamma and delta represent the factors used to correct the integrated peak areas of individual bilirubin photoisomers, and they are arranged in the order of the formula. It was demonstrated that the relative 455 nm molar absorptivity values for (ZZ)-bilirubin and all its geometrical and structural photoisomers, i.e., (ZZ)-bilirubin, (ZE)-bilirubin (EZ)-bilirubin, (EZ)-cyclobilirubin (= lumirubin) and (EE)-cyclobilirubin in the HPLC eluent, are, respectively, 1.0, 0.81 (= alpha), 0.54 (= beta), 0.47 (= gamma) and 0.39 (= delta).