An electrospray ionization source for thermochemical investigation with the guided ion beam mass spectrometer |
| |
Authors: | R. M. Moision P. B. Armentrout |
| |
Affiliation: | Department of Chemistry, University of Utah, Salt Lake City, Utah 84112, USA. |
| |
Abstract: | An electrospray ionization (ESI) source developed for use with the guided ion beam tandem mass spectrometer (GIBMS) is described. For accurate determination of thermochemistry using threshold collision-induced dissociation (TCID), it is essential that any source produces ions with four exacting characteristics: (1) high intensity, (2) stable signal, and well-defined energies both (3) kinetic, and (4) internal. To accomplish these objectives, the ions generated by the electrospray are collected using a radio frequency electrodynamic ion funnel and are then transferred into a hexapole ion guide where they are thermalized and subsequently passed into higher-vacuum regions for analysis. The resulting ion intensities using this source can exceed 10(6) ions/s. Stable beams (<10% variation in signal) can be generated over multiple hours. The kinetic energy distribution of ions emerging from this source has been shown to be well described by a Gaussian distribution with a full width half maximum (FWHM) of about 0.1-0.2 eV in the laboratory frame of reference. Finally, TCID results for ions generated with this source show excellent agreement with previously reported threshold values for ions generated using a variety of sources and experimental methodologies. This confirms that internal energies of the ions are well described by a Maxwell-Boltzmann distribution at room temperature. |
| |
Keywords: | |
本文献已被 ScienceDirect PubMed SpringerLink 等数据库收录! |
|