首页 | 本学科首页   官方微博 | 高级检索  
     检索      


LC Analysis of Hexafluorophosphate on a Monolithic Column: Application to the Analysis of Ionic Liquids
Authors:Ling Yang  Hong Yu  Siwen Li  Shuang Zhou
Institution:1. College of Chemistry and Chemical Engineering, Harbin Normal University, 150025, Harbin, People’s Republic of China
Abstract:A silica-based monolithic column with tetrabutylammonium hydroxide–phthalic acid–acetonitrile as mobile phase has been used for LC analysis of hexafluorophosphate with direct conductivity detection. The effects of mobile phase composition, column temperature, and flow rate on retention of hexafluorophosphate were investigated and the optimum chromatographic conditions were selected. Common anions and tetrafluoroborate did not interfere with analysis of hexafluorophosphate. The detection limit (S/N = 3) for hexafluorophosphate was 6.2 mg L?1. Relative standard deviation (n = 5) for peak area was 0.1%. The method was successfully used for analysis of hexafluorophosphate in ionic liquids.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号