首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Selecting parameters and limits for equipment operational qualification
Authors:L Huber  Lou Welebob
Institution:(1) Hewlett-Packard, D-76337 Waldbronn, Germany Tel.: +49-7243-602209; Fax: +49-7243-602501; e-mail: ludwig_huber@hp.com, DE;(2) Hewlett-Packard, 371 Hoes Lane, Piscataway, NJ 08854, USA Tel.: +1-980-562-5246; Fax: +1-908-562-6246; e-mail: lou_welebob@hp.com, US
Abstract:While operational qualification (OQ) is a well-established term within equipment qualification, users of equipment often become unsure when it comes to implementation. The biggest problem is how to select procedures and acceptance criteria. Should these be the vendor's specifications or should the users define their own limits, and, if so, how? Should all instruments of the same type have the same values or should these be optimized for each individual instrument? This article will provide an overall strategy and specific examples for HPLC on how to select procedures and acceptance limits that are based on efficient use of resources, on practicality and on the intended use of the equipment.
Keywords:  Equipment qualification Operational qualification  Implementation  Acceptance limits  Procedure selection  Abbreviations AU Absorption Units  DQ Design qualification  IQ Installation qualification  OQ Operational qualification  PQ Performance qualification  RSD Relative standard deviation  RT Retention time
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏