Determination of volatile bases in seafood using the ammonia ion selective electrode: collaborative study |
| |
Authors: | Ellis P C Pivarnik L F Thiam M Ellis P C Pivarnik L F Thiam M |
| |
Affiliation: | Rhode Island Department of Health Laboratories, Providence 02904, USA. |
| |
Abstract: | Nine collaborating laboratories tested a combination of 23 seafood samples for volatile bases using an ammonia ion selective electrode. Results were reported as mg NH3/100 g fish, but the method reflected levels of both ammonia and trimethylamine, which permeated the ammonia membrane. The 23 samples were broken down into 8 blind duplicate pairs, 2 Youden matched pairs, and 3 single samples covering fresh to spoiled product ranging from 8 to 82 mg NH3/100 g. Seven species were evaluated: Atlantic cod, squid, Atlantic halibut, gray sole, monkfish, dogfish, and Atlantic mackerel. The ammonia electrode assay was performed on an aqueous homogenate consisting of 95 mL distilled water and 5.0 g sample tissue. Alkaline ion strength adjusting solution (2 mL) was added to the homogenate to liberate ammonia that was sensed by the ion specific electrode and measured on a precalibrated portable meter. Repeatability standard deviations (RSDr) ranged from 4.2 to 17%; reproducibility standard deviations (RSDR) ranged from 8.8 to 21%. A standard ammonium chloride solution was provided to all laboratories to spike 3 different samples at 10 mg NH3/100 g. Recoveries of added ammonia as ammonium chloride for fresh, borderline, and spoiled samples were 88.6, 107, and 128%, respectively. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|