Determination of triazine herbicide residues in water samples by on-line sweeping concentration in micellar electrokinetic chromatography |
| |
Authors: | Shuai Hua Zhang Yuan Yuan Yang Dan Dan Han Chun Wang Xin Zhou Xiao Huan Zang Zhi Wang |
| |
Affiliation: | Key Laboratory of Bioinorganic Chemistry, College of Sciences, Agricultural University of Hebei, Baoding 071001, China |
| |
Abstract: | A new method for the determination of atrazine, simazine and prometryn in water samples by on-line sweeping concentration technique in micellar electrokinetic chromatography (MEKC) was developed. Various parameters affecting sample enrichment and separation efficiency were systematically studied. Compared with the conventional MEKC method, up to 60-200-fold improve-merit in concentration sensitivity was achieved in terms of peak height by using this sweeping injection technique. The compound strychnine was used as the internal standard for the improvement of the experimental reproducibility. The limits of detection (S/ N = 3:1) for atrazine, simazine and prometryn were 9, 10 and 0.5 ng mL-1, respectively. This method has been successfully applied to the analysis of atrazine, simazine and prometryn in lake, steam and ground water. |
| |
Keywords: | Triazine herbicides On-line concentration MEKC Sweeping Water samples |
本文献已被 维普 万方数据 ScienceDirect 等数据库收录! |
| 点击此处可从《中国化学快报》浏览原始摘要信息 |
|
点击此处可从《中国化学快报》下载全文 |
|