首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Direct solid vs. slurry analysis of tobacco leaves for some trace metals by graphite furnace AAS
Authors:R Dobrowolski and J Mierzwa
Institution:(1) Central Laboratory, University of M. Curie-Sklodowska, PL-20-031 Lublin, Poland
Abstract:Summary Graphite furnace atomic absorption spectrometry (GFAAS) is used to determine cadmium, lead, nickel and cobalt in two samples of tobacco leaves (candidate reference materials). Two techniques for the direct determination of these elements are investigated — direct solid sampling into a ldquoring chamber tuberdquo specially designed for the Zeiss AAS-3 spectrometer and slurry sampling. For both investigated sampling methods the optimum parameters of temperature-time programs, influence of Pd modifier and calibration methods are discussed. There is acceptable agreement between the results obtained from direct solid, slurry and solution (after wet decomposition) sampling.Presented at the 4th International Colloquium on Solid Sampling with Atomic Spectroscopy, October 8–10, 1990; Jülich, Federal Republic of Germany
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏