Comparison of three analytical methods for quality control of clozapine tablets |
| |
Authors: | Raggi M A Pucci V Bugamelli F Volterra V |
| |
Institution: | University of Bologna, Department of Pharmaceutical Sciences, Italy. |
| |
Abstract: | Three different analytical methods for the quality control of clozapine in commercial formulations were developed and compared: a liquid chromatographic (LC) method with UV detection, a capillary zone electrophoretic (CZE) method, and a linear scan voltammetric (LSV) method. The isocratic LC procedure used a C18 reversed-phase column; the CZE method used an uncoated fused-silica capillary and phosphate buffer containing polyvinylpyrrolidone as the background electrolyte; the LSV method analyzed clozapine solutions with acidic phosphate buffer as the supporting electrolyte. The 3 methods gave similar and satisfactory results, in terms of precision and accuracy. Repeatability and intermediate precision were good (RSD% < 2.2) and accuracy, resulting from recovery studies, was between 98 and 102%. The rapidity of analysis was high for all 3 methods, especially for the LSV. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|